Item
Air Velocity Transmitter
Air Velocity Accuracy
+/-3% (32 to 122 Degrees F)
Velocity Ranges
FPM: 250, 500, 1000, 2000, 3000, 5000, 10000, 15000
Velocity Accuracy
3% FS Process gas: 32 to 122 Degrees F (0 to 50 Degrees C); 4% FS Process gas: -40 to 32 Degrees F u0026 122 to 212 Degrees F (-40 to 0 Degrees C u0026 50 to 100 Degrees C)
Power Required
10 to 16 VAC 1.5A Rating Required on Supply Due to Initial Power Surge Drawn by Transmitter, 12 to 35 VDC
Response Time (Sec.)
1.5 Seconds to 95% of Final Value (Output Filter Set to Minimum)
Max. Pressure (PSI)
100 psi
Operating Temp.
Process: -40 to 212 Degrees F (-40 to 100 Degrees C) Ambient: 32 to 140 Degrees F (0 to 60 Degrees C)
Manufacturers Warranty Length
1
Maximum Feet Per Minute
15000 FPM
Maximum DC Voltage Required
35 V
Minimum AC Voltage Required
10 V
Minimum DC Voltage Required
12 V
Maximum Output Signal
20 mA
Maximum Velocity Measured
15,000 fpm
Minimum Operating Temperature
32 Degrees F
Manufacturer Warranty Length
1 yr
Digital Display Screen Type
LED
Maximum Operating Temperature
140 Degrees F
Minimum Temperature Measured
-40 Degrees F
Maximum AC Voltage Required
16 V
Maximum Probe Length
24 in
Minimum Velocity Measured
250 fpm
Maximum Temperature Measured
212 Degrees F
Minimum Output Signal
4 mA
Country of Origin (subject to change)
United States